Accuracy
|
±2 % plus 0.01 NTU from 0 – 40 NTU;
±10 % of reading from 40 – 1000 NTU based on
Formazin primary standard (at 25 °C)
|
Calibration Method
|
For Formazin and Stablcal:
20 NTU from 0 to 40 NTU; at 20 FNU and 600 NTU for full range
Custom calibration for up to 6-point calibrations
Degrees 0 – 100 mg/L ; SDVB 0 – 1000 NTU
|
Certifications
|
CE compliant
US FDA accession number:
1420493-000 EPA version, 1420492-000ISO version
Complies with IEC/EN 60825-1 and to 21 CFR 1040.10 in accordance with Laser Notice No. 50) ; Australian ACMA Marking
|
Communication
|
Link2sc and RFID
|
Dimensions
(H x W x D)
|
105 mm x 409 mm x 278 mm
|
Display
|
7″ WVGA (800 x 480) color TFT-LCD with touch screen
|
Enclosure Construction
|
PC/ABS Blend Plastic
|
Light Source
|
Class 2 laser product, with embedded 650 nm (EPA 0.43 mW) or
Class 1 laser product, with embedded 850 nm (ISO),
max. 0.55 mW (complies with IEC/EN 60825-1 and to 21 CFR 1040.10 in accordance with Laser Notice No. 50)
|
Model
|
TU5200 EPA
|
Operating Humidity
|
80 % at 30 °C (non condensing)
|
Operating Temperature Range
|
10 – 40 °C
|
Power Requirements (Hz)
|
50/60 Hz
|
Power Requirements (Voltage)
|
100 – 240 VAC
|
Range
|
EPA:
0 – 700 NTU / FNU / TE/F / FTU
0 – 100 mg/L ; 0 – 175 EBC
ISO:
0 – 1000 NTU / FNU / TE/F / FTU
0 – 100 mg/L ; 0 – 250 EBC
|
Regulatory
|
EPA
|
Repeatability
|
<40 NTU: Better than 1% of reading or ±0.002 NTU on Formazin at 25 °C (77 °F), whichever is greater
>40 NTU: Better than 3.5% of reading on Formazin at 25 °C (77 °F)
|
Resolution
|
0.0001 NTU / FNU / TE/F / FTU / EBC / mg/L
|
Response Time
|
Signal averaging off: 7 seconds
Signal averaging on: 10 seconds
(when averaging time is 5 seconds)
|
Sample Temperature
|
4 – 70°C (39 – 158°F)
|
Storage Conditions
|
-30 to 60 °C (-22 to 140 °F)
|
Stray Light
|
<10 mNTU
|
Units
|
NTU, FNU, TE/F, FTU, EBC;
mg/L if calibrated with Degrees calibration curve
|
Warranty
|
1 YEAR
|
Weight
|
2.4 kg
|