Catalog Number |
GENESYS 40 Visible Spectrophotometer |
840-297000 |
|
Optical Design |
Dual beam |
Spectral Bandwidth |
5 nm |
Wavelength Range |
325 nm – 1100 nm |
Detector Type |
Dual Silicon Photodiodes |
Display |
7 – Inch color touchscreen, fixed, high definition, 800 × 1280 pixels |
PhotometricAccuracy Instrument |
± 0.002A at 0.5A ; ± 0.004A at 1.0A ; ±0.008A at 2.0A |
Printer |
Snap-on printer available |
Wavelength Accuracy |
± 0.5 nm |
Baseline Flatness |
± 0.005A |
Compartments |
|
Connections |
|
Description |
GENESYS 40 Visible Spectrophotometer |
Dimensions(L x W x H) |
35.5 x 38.5 x 19.5 cm |
Drift |
(At 500 nm after 1 hour warmup) <0.0010A/Hr |
ElectricalRequirements |
External AC to DC converter. Voltage and Frequency (Hz)selected automatically, 100 – 240 Volts, 50 – 60 Hz. |
Keypad |
Touchscreen |
Lamp |
Tungsten-Halogen (>1000 hours) |
Noise |
(RMS at 500 nm 60 consecutive measurements) :
|
Optical Design |
Dual Beam |
Photometric Display |
-3A to +5A |
Photometric Range |
–3A to +3.5A |
PhotometricRepeatability |
±0.001A at 1A measured at 1.0A at 546 nm |
Spectral Bandwidth |
5 nm |
Stray Light |
<0.05%T at 340 nm and 400 nm |
Wavelength Data Interval |
0.2 nm, 0.5 nm, 1 nm, 2 nm, 5 nm |
Wavelength Range |
325 nm – 1100 nm |
Wavelength Repeatability |
< ±0.2 nm |
Wavelength Scan Speed |
Automatic – up to 1800 nm/min |
Weight (Metric) |
7.5 kg |
Unit Size |
Each |
Catalog Number |
GENESYS 50 UV-Vis Spectrophotometer |
840-298000 |
|
Optical Design |
Dual Beam |
Spectral Bandwidth |
2 nm |
Wavelength Range |
190 nm – 1100 nm |
Detector Type |
Dual Silicon Photodiodes |
Display |
7 – inch color touchscreen, fixed, high definition, 800 × 1280 pixels |
Photometric
|
±0.002A at 0.5A ±0.004A at 1.0A ±0.008A at 2.0A |
Printer |
Snap-on printer available |
Wavelength Accuracy |
±0.5 nm |
Baseline Flatness |
±0.002A |
Compartments |
|
Connections |
|
Description |
GENESYS 50 UV-Vis Spectrophotometer |
Dimensions (L x W x H) |
35.5 x 38.5 x 19.5 cm |
Drift |
(At 500 nm after 1 hour warmup) <0.0005A/Hr |
ElectricalRequirements |
External AC to DC converter. Voltage and Frequency (Hz)selected automatically, 100 – 240 Volts, 50 – 60 Hz. |
Keypad |
Touchscreen |
Lamp |
Xenon Flash Lamp (>5 years typical, 3 years guaranteed) |
Noise |
(RMS at 500 nm 60 consecutive measurements)
|
Optical Design |
Dual beam |
Photometric Display |
-3A to +5A |
Photometric Range |
–2A to +3.5A |
Photometric Repeatability |
±0.001A at 1A measured at 1.0A at 546 nm |
Spectral Bandwidth |
2 nm |
Stray Light |
|
Wavelength Data Interval |
0.2 nm, 0.5 nm, 1 nm, 2 nm, 5 nm |
Wavelength Range |
190 nm – 1100 nm |
Wavelength Repeatability |
< ±0.2 nm |
Wavelength Scan Speed |
Slow, medium and fast – up to 1600 nm/min |
Weight (Metric) |
7.5 kg |
Unit Size |
Each |
Catalog Number |
GENESYS 30 Visible Spectrophotometer |
840-277000 |
|
Detector Type |
Silicon Photodiode |
Display |
5 in. diagonal, 32 bit color display, 800 x 480 pixels |
PhotometricAccuracy Instrument |
± 0.002A (0 – 0.3A) 0.5 % of ABS reading (0.301A – 2.5A) |
Printer |
Thermal printer accessory with 80mm width paper |
Wavelength Accuracy |
± 2nm |
Baseline Flatness |
< 0.003A 325nm – 1000nm smoothed |
Compartments |
|
Connections |
|
Description |
GENESYS 30 Visible Spectrophotometer |
Dimensions (L x W x H) |
35.5 cm x 38.5 cm x 19.5 cm |
Drift |
< 0.002A/Hr At 500nm after 2hr warmup.Typical < 0.004A/hr after 1hr warmup |
ElectricalRequirements |
External AC to DC converter.Voltage and Hz selected automatically, 100 – 250 Volts, 50 – 60 Hz |
Keypad |
Tactile rubber 23 keys with numeric keypad |
Lamp |
1000 Hrs typical service life |
Languages |
English, Spanish, German, French, Italian, Portuguese, Russian,Chinese, Japanese, Korean, Thai, Arabic |
Noise |
RMS at 500 nm. 60 consecutive measurements |
Optical Design |
Single beam |
Photometric Display |
-3A to +3A, 0 to 200 000 %T, 0 to 9 999 999 C |
Photometric Range |
-3A to +3A |
Photometric Repeatability |
± 0.002 A Measured at 1.0 A @546nm |
Spectral Bandwidth |
5 nm |
Stray Light |
< 0.1%T at 340 nm and 400 nm Measured using glasscutoff filters or NaNO2 at 340 nm |
Wavelength Data Interval |
1nm |
Wavelength Range |
325 nm – 1100 nm |
Wavelength Repeatability |
< ±1nm |
Wavelength Scan Speed |
Automatic – up to 1200 nm/min |
Weight (Metric) |
7.5 kg |
Unit Size |
Each |