Automatic dead weight hardness tester for micro and macro Vickers and Knoop method, single or multiple indentations. DM8 and DM2 are low cost microhardness testers with high accuracy standards. They are equipped with a motorized rotating turret with combined magnification lenses and indenters for reading and indenting. For micro and macro Vickers and case depth test on all metals: Iron, steel, tempered steel, cast iron, brass, aluminum, copper and metal alloys. Heat treatment, hardening, nitriding, cementation and hardfacing. Knoop test on ceramic and glass materials.

Test Loads : from 1 gf to 2 kgf (from 0,0098 to 19,614 N)

Vickers ISO 4545 / ASTM E-384 :

HV0,001 – HV0,003 – HV0,005 – HV0,01 – HV0,025 – HV0,05 – HV0,1 – HV0,2 – HV0,5 – HV1 – HV2

Knoop ISO 6507 / ASTM E-384 :

HK0,01 – HK0,025 – HK0,05 – HK0,1 – HK0,2 – HK0,5 – HK1

 

TECHNICAL FEATURES

DM2

Auto/Semi/Manual

DM8 A

Auto/Semi/Manual

DM8 B

Auto/Semi/Manual

Standards

DIN-EN-ISO 6507 / ISO 4545 / ASTM E-384 / JIS B 7734

X/Y Table

Manual / Semi: Manual table 100 x 100 mm with 10 μm step

Auto: Motorized table 100 x 60 mm or 200 x 100 mm (automatic) with 0.5 μm step

Elevating screw stroke

0/110 mm

0/120 mm

Depth capacity

0/70 mm

0/160 mm

Turret (Indenter/Objectives)

Motorized – 4 slots

Motorized – 6 slots

Indent reading

Auto: Automatic focus with auto reading

Semi: Manual focus with auto reading

Manual: Manual focus and manual reading

Test loads

0.09807 – 0.2452 – 0.4903 – 0.9807 – 1.961 – 2.942 – 4.903 – 9.807 N

0.04903 – 0.09807 – 0.2452 – 0.4903 – 0.9807 – 1.961 – 2.942 – 4.903 – 9.807 N

0.0098 – 0.0294 – 0.04903 – 0.09807 – 0.2452 – 0.4903 – 0.9807 – 1.961 – 2.942 – 4.903 – 9.807 – 19.614 N

(10 – 25 – 50 gf / 0.1 – 0.2 – 0.3 – 0.5 – 1 kgf)

(5 – 10 – 25 – 50 gf / 0.1 – 0.2 – 0.3 – 0.5 – 1 kgf)

(1 – 3 – 5 – 10 – 25 – 50 gf / 0.1 – 0.2 – 0.3 – 0.5 – 1 – 2 kgf)

Feasible tests

Vickers (Knoop on request)

Accuracy

Better than 1 %