Item |
Specifications |
|
GeneralParameters |
Power Supply |
AC220V, 50/60 Hz (Standard)1320VA ; 110V 60Hz (Optional) |
Size (L*W*H) |
950*645*(1250-11450) mm |
|
Maximum heater output power (220VAC) |
Air chamber heater: 330W; |
|
Water tank heater: 155W |
||
Auxillary network power output |
AC 220V / 50Hz, Both maximum permissible currents are 220V 1.5A |
|
Heater indication |
0-100%, 10% increasing amount |
|
Temperature control mode |
Bed temperature mode ; The skin temperature mode |
|
Bed temperature control |
25.0 – 37.0°C; |
|
temperature range |
37.1 – 39.0°C (Span Mode) |
|
Skin temperature sensor temperature display range |
5.0 – 65.0°C |
|
Heating time |
< 50min |
|
Temperature variability |
< 0.5°C |
|
Temperature uniformity(mattress in horizontal position) |
< 0.8°C |
|
Temperature uniformity(mattress in tilted position) |
< 1.0°C |
|
Difference between average incubator temperature and control temperature |
< 1.5°C |
|
Skin temperature sensor accuracy |
=+0.3°C’ |
|
Humidity
|
The running time of the tank after it is filled with distilled water |
Humidity below 70%RH, at least 24h |
Water tank capacity |
1500ml |
|
Humidity display range |
0%RH – 99%RH |
|
Humidity control range |
CAH%RH – 90%RH(1% per adjustment) |
|
Humidity control accuracy |
1 |
|
Oxygen
|
Oxygen concentrationdisplay range |
0% – 99% |
Oxygen concentration display resolution |
1% |
|
Oxygen concentration control accuracy |
=+5% oxygen volume concentration’ |
|
Oxygen concentration setting range |
21% – 60% (1% per adjustment) |
|
Oxygen sensorservice life |
The maximum use time is 10000h at100 % oxygen concentration |
|
Transport
|
Temperature |
-2°C – +55°C |
Atmospheric pressure range |
500hPa – 1060hPa |
|
Relative humidity |
< 93% RH |
|
Packing Size /
|
Base packing size: 1200*800*870mm 50kg |
|
Body packing size: 1100*710*855mm 60kg |
||
The
|
Relative humidity |
30%RH – 75%RH |
The environment temperature |
=+20°C – + 30°C’ |
|
Pressure range |
700hPa – 1060hPa |
|
Service altitude |
< 2000m |
|
Over voltage category |
II |
|
Pollution levels |
2 |
|
Ambient air velocity |
< 0.3m/s |
|
Other
|
Environmental noise inincubator baby compartment |
No more than 55dB(A) at stable temperature |
Carbon dioxide concentration in the infant compartment |
A mixture of 4% carbon dioxide and air was fed into an 8mm diameter tube in the vertical direction from the mat to the top of the box at 750mL/min to a point 10cm above the mattress center and 10cm above the center. The concentration was less than 0.5% after stabilization |
|
The velocity of air flow over a baby’s mattress |
< 0.35m/s |
|
GeneralAlarm |
Power off alarm, Fan alarm, sensor alarm, Deviation alarm, Over temperature alarm, Water tank placement error alarm, Water shortage alarm |
|
TheSystemAlarm |
Mcu1 and Mcu2 communication error alarm, Mcu2 and Mcu3 communication error alarm, Analog-to-digital conversion chip 2544 fault alarm, Analog-to-digital conversion chip 2548 fault alarm, Over temperature alarm, Temperature heating system fault, Humidity heating system fault |
Configuration : |
||
Model |
BK-3201 (I) |
BK-3201 (II) |
Temp.Control |
Bed/Skin temperature |
Bed/Skin temperature |
Display |
LCD |
LCD |
Bed Temp. Control Range |
25°C – 39°C |
25°C – 39°C |
Skin Temp. Control Range |
34°C – 38°C |
/ |
Humidity Control Range |
CAH% – 99%RH |
CAH% – 99%RH |
Oxygen Control Range |
21% – 99% |
/ |
Cabinet |
Yes |
Yes |