XP Analytical Balances |
||||
Technical Data (Limit Values) |
XP105DR |
XP205 |
XP205DR |
|
Maximum Capacity |
120 g |
220 g |
220 g |
|
Maximum Capacity, Fine Range |
31 g |
— |
81 g |
|
Readability |
0.1 mg |
0.01 mg |
0.1 mg |
|
Readability, Fine Range |
0.01 mg |
— |
0.01 mg |
|
Repeatability |
– (at nominal load) |
0.06 mg (100 g) |
0.03 mg (200 g) |
0.06 mg (200 g) |
– (at low load) |
0.05 mg (10 g) |
0.015 mg (10 g) |
0.05 mg (10 g) |
|
– (at low load),
|
0.015 mg (10 g) |
— |
0.015 mg (10 g) |
|
Linearity |
0.15 mg |
0.1 mg |
0.15 mg |
|
Excentric Load Deviation (test load) 1) |
0.2 mg (50 g) |
0.2 mg (100 g) |
0.25 mg (100 g) |
|
Sensitivity Offset |
4×10-6 Rnt |
2×10-6 Rnt |
2.5×10-6 Rnt |
|
Sensitivity Temperature Drift 2) |
1×10-6/°C Rnt |
1×10-6/°C Rnt |
1×10-6/°C Rnt |
|
Sensitivity Stability 3) |
1×10-6/a Rnt |
1×10-6/a Rnt |
1×10-6/a Rnt |
|
Interface Update Rate |
23/s |
23/s |
23/s |
|
Click Here : METTLER TOLEDO [XP Analytical Balance & XP Mricrobalance] Brochure |
||||
XP Analytical Balances |
||||
Technical Data (Limit Values) |
XP204 |
XP504 |
XP504DR |
|
Maximum Capacity |
220 g |
520 g |
520 g |
|
Maximum Capacity, Fine Range |
— |
— |
101 g |
|
Readability |
0.1 mg |
0.1 mg |
1 mg |
|
Readability, Fine Range |
— |
— |
0.1 mg |
|
Repeatability |
– (at nominal load) |
0.07 mg (200 g) |
0.12 mg (500 g) |
0.6 mg (500 g) |
– (at low load) |
0.05 mg (10 g) |
0.1 mg (10 g) |
0.5 mg (10 g) |
|
– (at low load),
|
— |
— |
0.1 mg (10 g) |
|
Linearity |
0.2 mg |
0.4 mg |
0.5 mg |
|
Excentric Load Deviation (test load) 1) |
0.25 mg (100 g) |
0.4 mg (200 g) |
0.5 mg (200 g) |
|
Sensitivity Offset |
3×10-6 Rnt |
3×10-6 Rnt |
4×10-6 Rnt |
|
Sensitivity Temperature Drift 2) |
1×10-6/°C Rnt |
1×10-6/°C Rnt |
1×10-6/°C Rnt |
|
Sensitivity Stability 3) |
1×10-6/a Rnt |
1×10-6/a Rnt |
1×10-6/a Rnt |
|
Interface Update Rate |
23/s |
23/s |
23/s |