Infant Radiant Warmer
|
Item
|
Specifications
|
General
Parameters
|
Power Supply
|
AC 220V, 50/60 Hz (Standard)
1320 VA; 110V 60Hz (Optional)
|
Size (L*W*H)
|
950 * 645 * (1250 – 11450) mm
|
Maximum Heater
Output Power (220VAC)
|
Air chamber heater: 330W;
Water tank heater: 155W
|
Auxillary network
power output
|
AC 220V / 50Hz, Both maximum permissible
currents are 220V 1.5A
|
Heater indication
|
0-100%, 10% increasing amount
|
Temperature control mode
|
Bed temperature mode ; The skin temperature mode
|
Bed temperature control
|
25.0 – 37.0°C;
|
temperature range
|
37.1 – 39.0°C (Span Mode)
|
Skin temperature sensor
temperature display range
|
5.0 – 65.0°C
|
Heating time
|
< 50min
|
Temperature variability
|
< 0.5°C
|
Temperature uniformity
(mattress in horizontal position)
|
< 0.8°C
|
Temperature uniformity
(mattress in tilted position)
|
< 1.0°C
|
Difference between average
incubator temperature and control temperature
|
< 1.5°C
|
Skin temperature sensor accuracy
|
+0.3°C
|
Humidity
Control
|
The running time of the tank
after it is filled with distilled water
|
Humidity below 70 % RH,
at least 24h
|
Water tank capacity
|
1500ml
|
Humidity display range
|
0% RH – 99% RH
|
Humidity control range
|
CAH % RH – 90 % RH
(1% per adjustment)
|
Humidity control accuracy
|
1
|
Oxygen
Concentration
Control
|
Oxygen concentration display range
|
0% – 99%
|
Oxygen concentration display resolution
|
1%
|
Oxygen concentration control accuracy
|
+5% oxygen volume concentration
|
Oxygen concentration setting range
|
21% – 60% (1% per adjustment)
|
Oxygen sensor
service life
|
The maximum use time is 10000 h at
100 % oxygen concentration
|
Transport
Storage
|
Temperature
|
-2°C – +55°C
|
Atmospheric pressure range
|
500hPa – 1060hPa
|
Relative humidity
|
< 93% RH
|
Packing size / Gross weight
|
Base packing size : 1200 * 800 * 870 mm; 50kg
|
Body packing size: 1100 * 710 * 855 mm; 60kg
|
The Work
Environmental
|
Relative humidity
|
30%RH – 75%RH
|
The environment temperature
|
“+20°C – + 30°C
|
Pressure range
|
700hPa – 1060hPa
|
Service altitude
|
< 2000m
|
Over voltage category
|
II
|
Pollution levels
|
2
|
Ambient air velocity
|
< 0.3m/s
|
General Alarm
|
-
Environmental noise in incubator baby compartment
-
Carbon dioxide concentration in the infant compartment
-
The velocity of air flow over a baby’s mattress
-
Power off alarm, Fan alarm, sensor alarm, Deviation alarm, Over temperature alarm, Water tank placement error alarm, Water shortage alarm
|
-
No more than 55dB(A) at stable
-
temperature
-
A mixture of 4% carbon dioxide and air was fed into an 8 mm diameter tube in the vertical direction from the mat to the top of the box at 750 mL / min to a point 10 cm above the mattress center and 10 cm above the center.
-
The concentration was less than 0.5% after stabilization < 0.35m/s
-
Power off alarm, Fan alarm, Sensor alarm, Deviation alarm, Over temperature alarm, Water tank placement error alarm, Water shortage alarm
|
The System
Alarm
|
MCU1 and MCU2 communication error alarm, MCU2 and MCU3 communication error alarm,Analog-to-digital conversion chip 2544 fault alarm, Analog-to-digital conversion chip 2548 fault alarm, Over temperature alarm, Temperature heating system fault,Humidity heating system fault
|