Catalog Number |
GENESYS™ 30 Visible Spectrophotometer |
840-277000 |
|
Connections |
::) USB-A on front face supports flash memory devices for method and data storage::) USB-B on side supports connection to a Windows™ computer running optional remote control software. |
Detector Type |
Silicon Photodiode |
Display |
5 in. diagonal, 32 bit color display, 800 x 480 pixels |
Photometric Accuracy Instrument |
±0.002A (0 – 0.3A) ; 0.5% of ABS reading (0.301A – 2.5A) |
Printer |
Thermal printer accessory with 80mm width paper |
Wavelength Accuracy |
± 2nm |
Wavelength Range |
325 nm – 1100 nm |
Baseline Flatness |
<0.003A ; 325nm – 1000nm smoothed |
Compartments |
|
Description |
GENESYS 30 Visible Spectrophotometer |
Dimensions (L x W x H) |
35.5cm x 38.5cm x 19.5cm |
Drift |
<0.002A/HrAt 500nm after 2hr warmup.Typical <0.004A/hr after 1hr warmup |
ElectricalRequirements |
External AC to DC converter.Voltage and Hz selected automatically, 100-250 volts, 50-60Hz |
Keypad |
Tactile rubber 23 keys with numeric keypad |
Lamp |
1000 Hrs typical service life |
Languages |
English, Spanish, German, French, Italian, Portuguese, Russian,Chinese, Japanese, Korean, Thai, Arabic |
Noise |
≤0.001A at 0A≤0.001A at 1A≤0.002A at 2ARMS at 500nm. 60 consecutive measurements |
Optical Design |
Single beam |
Photometric Display |
-3A to +3A, 0 to 200 000 %T, 0 to 9 999 999 C |
Photometric Range |
-3A to +3A |
Photometric Repeatability |
± 0.002 A ; Measured at 1.0 A @546nm |
Spectral Bandwidth |
5 nm |
Stray Light |
<0.1%T at 340nm and 400nmMeasured using glass cutoff filters or NaNO2 at 340nm |
Wavelength Data Interval |
1nm |
Wavelength Repeatability |
< ±1nm |
Wavelength Scan Speed |
Automatic – up to 1200nm/min |
Weight (Metric) |
7.5kg |
Warranty |
1 Year |
Unit Size |
Each |
Description |
GENESYS 30 |
GENESYS 40 |
GENESYS 140 |
GENESYS 50 |
GENESYS 150 |
BioMate 160 |
GENESYS 180 |
|
Optical Design |
Single Beam |
Dual Beam |
Double Beam |
|||||
Spectral Bandwidth |
5 nm |
2 nm |
||||||
Light Source(Typical Lifetime) |
Tungsten-Halogen (>1000 hours) |
Xenon Flash Lamp (>5 years typical, 3 years guaranteed) |
||||||
Detector |
Silicon Photodiode |
Dual Silicon Photodiodes |
||||||
Wavelength |
Range |
325–1100 nm |
190–1100 nm |
|||||
Accuracy |
±2 nm |
±0.5 nm |
||||||
Repeatability |
<±1 nm |
<±0.2 nm |
||||||
ScanSpeed |
Automatic – up to1200 nm/min |
Automatic – up to 1800 nm/min |
Slow, medium and fast (up to 1600 nm/min) |
|||||
DataResolution |
1 nm |
0.2 nm, 0.5 nm, 1 nm, 2 nm, 5 nm |
0.1 nm, 0.2 nm, 0.5 nm,1 nm, 2 nm, 5 nm |
|||||
Photometric |
Range |
–3A to +3A |
–3A to +3.5A |
–2A to +3.5A |
||||
Display |
–3A to +3A, 0 to 200 000 %T,0 to 9 999 999 C |
–3A to +5A |
||||||
Accuracy |
± 0.002A (0 – 0.3A) 0.5 % ofABS reading (0.301A – 2.5A) |
± 0.002A at 0.5A± 0.004A at 1.0A± 0.008A at 2.0A |
||||||
Repeatability1 |
± 0.002 A |
± 0.001A at 1A |
||||||
Noise2 |
≤ 0.001A at 0A≤ 0.001A at 1A≤ 0.002A at 2A |
≤0.00020A at 0A at 260 and 500 nm≤0.00030A at 1A at 260 and 500 nm≤0.00040A at 2A at 260 and 500 nm |
||||||
Drift3 |
< 0.002A/Hr |
< 0.0010A/Hr |
<0.0005A/Hr |
|||||
Stray Light |
< 0.1%T at 340 nm and 400 nm |
< 0.05%T at 340 nm and 400 nm |
< 1.0 %T 198 nm (KCl) , < 0.05 %T at 220 nm (NaI),< 0.03 %T at 340 nm (NaNO2) |
|||||
Baseline Flatness |
< 0.003A |
± 0.005A |
± 0.002A |
|||||
Display |
5-inch diagonal, 32-bit color display,800 × 480 pixels |
7-inch color touchscreen,fixed, high definition,800 × 1280 pixels |
7-inch color touchscreen,tiltable, high definition,800 × 1280 pixels |
7-inch color touchscreen,fixed, high definition,800 × 1280 pixels |
7-inch color touchscreen, tiltable,high definition, 800 × 1280 pixels |
|||
Keypad |
Tactile rubber 23 keys with numeric keypad |
Touchscreen |
||||||
SampleCompartment |
• Accessible from top, front or side• All platforms can accommodate 100 mm cells• Available accessory for test-tubes up to 25 mmdiameter and 150 mm height• Removable, washable sample compartment linerwith magnetic placement and hold-down |
Accommodates (optional):• 8-position cell changer• 4-position cell changer(long path cells)• Peltier thermostattedcell holder (20–60 °C)• Sipper accessory |
• Accessible from top,front or side• Able to accommodatecells up to 100 mm pathlength• Available accessory for test-tubesup to 25 mm diameter and150 mm height• Removable, washable samplecompartment liner with magneticplacement and hold-down |
Accommodates (optional):• 8-position cell changer• 4-position cell changer(long path cells)• Peltier thermostattedcell holder (20–60 °C)• Sipper accessory• Fiber optic probe coupler |
• 8-position cell changer(standard) Alsoaccommodates (optional):• 4-position cell changer(long path cells)• Peltier thermostattedcell holder (20–60 °C)• Sipper accessory• Fiber optic probe coupler |
|||
Printer |
Snap-on printer available |
|||||||
Connectivity |
• USB-A on front face supports flash memory devices for method and data storage• USB-B on side supports connection to a Windows computer running optional remote control software; export data to PC via USB |
• Single USB-A supports flash memory devices for method and data storage• Duplex USB-A supports connection to a Windows computer running optional remote control software, keyboard, mouse• Export data to network or PC via Ethernet or Wi-Fi• Print via USB, Ethernet or Wi-Fi |
||||||
Languages |
English (Available soon: Spanish, German, French, Italian, Portuguese, Russian, Chinese, Japanese, Korean, Thai) |
|||||||
Dimensions |
35.5 × 38.5 × 19.5 cm (L × W × H) |
|||||||
Weight |
7.5 kg |
|||||||
Power Requirements |
External AC to DC converter. Voltage and Frequency (Hz) selected automatically, 100–240 volts, 50–60 Hz. |
|||||||
Warranty |
1 (one) year |