BUHLER – Gas Cooler Series RC 1.1

BUHLER – Gas Cooler Series RC 1.1

 

BUHLER – Gas Cooler Series RC 1.1

 

Compact design: Pre-installed and ready to connect
Choose from one or two gas paths
Heat exchanger made from stainless steel, Duran glass and PVDF
Adjustable outlet dew point and alarm thresholds
Cooling block temperature display
Status display and output
Rated cooling power 360 kJ/h
Constant dew point stability ± 0.1 °C
Optional 4 – 20 mA or Modbus RTU signal output
Moisture detector, filter and condensate pump optional
Successor to the EGK 1/2 and EGK 1SD
Click Here : Sample Gas Cooler RC 1.1 Brochure
Click Here : Gas Cooler Series TC-MIDI Brochure

BUHLER – Gas Cooler Series RC 1.1

 

Gas Cooler Technical Data

​GAS COOLER SERIES RC 1.1

Ready for Operation

after max. 15 minutes

Rated Cooling Capacity (at 25°C)

360 kJ/h

Ambient Temperature

5 °C to 50 °C

Gas Outlet Dew Temperature,Preset

approx. 5 °C

Dew Point Fluctuations Static:

± 0.1 K

In the Entire Specification Range:

± 1.5 K

IP rating

IP 20

Housing

Stainless steel

Packaging Dimensions

Approx. 420 x 440 x 350 mm

Weight Incl. Heat Exchanger

approx. 15 kg

Electric Supply

115 V, 60 Hz or 230 V, 50/60 Hz ± 5% Plug per DIN EN 175301-803

Electrical Data

230 V 115 V

Typical Power Input :

296 VA 302 VA

Max. Operating Current :

2 A 3.5 A

Alarm Output Switching Connection

max. 250 V, 2 A, 50 VA Connector per DIN EN 175301-803

Installation

stand-alone or wall-mounted

Technical Data Peristaltic Pumps CPsingle / CPdouble

Flow Rate

0.3 L/h (50 Hz) / 0.36 L/h (60 Hz) with standard hose

Vacuum Inlet

max. 0.8 bar

Pressure Inlet

max. 1 bar

Outlet Pressure

1 bar

Hose

4 x 1.6 mm

Condensate Outlet

Hose nipple Ø6 mm Screw connection 4/6 (metric), 1/6″-1/4″ (US)

Protection Class :

IP 40

Materials Hose :

Norprene (Standard), Marprene,Fluran

Connections:

PVDF

Analogue Output Cooler Temperature Technical Data

Signal

4-20 mA or 2-10 Vcorresponds to -20 °C to +60 °C cooler temperature

Connection

M12x1 Plug, DIN EN 61076-2-101

Technical Data Filter AGF-PV-30-F2-L

Ambient Temperature

3 °C to 100 °C

Max. Operating Pressure with Filter

2 bar

Filter surface

125 cm2

Filter mesh

2 μm

Dead volume

108 ml

Materials

Filter:

PVDF, Duran Glass (parts in contact with media)

Seal

Viton

Filter element

sintered PTFE

Technical Data FF-3-N Moisture Detector

Ambient Temperature

3 °C to 50 °C

Max. Operating Pressure with FF-3N

2 bar

Material

PVDF, PTFE, epoxy resin, stainless steel 1.4571, 1.4576

Many gas analysis processes require extracting sample gas from the process. This also extracts process-related contamination such as particles or moisture. These can impact the measurement results or damage the measuring cells. The sample gas must therefore be conditioned before entering the analyser. The sample gas cooler reduces the gas temperature to below the dew point for this purpose, causing moisture to drop out, which is then discharged as condensate.

In addition to the status output to monitor the sample gas cooler function, we offer an optional 4 – 20 mA analog output or digital interface. The process control can access the process and diagnostic data via the Modbus RTU interface as well as configure the device settings.

The RC 1.1 is a compressor sample gas cooler designed for high cooling output and high ambient temperatures. Depending on the application, it can be configured with one or two gas paths. This makes it a vital cooling component in sophisticated analysis systems.

BUHLER – Gas Cooler Series RC 1.1

 

Heat Exchanger

TS TS-I2)

TG TG

TV-SS TVSS-I 2)

Version / Material

Stainless Steel

Glass

PVDF

Flow Rate vmax 1)

530 L/h

280 L/h

155 L/h

Inlet Dew Point τe,max1)

80 °C

80 °C

65 °C

Gas Inlet Temperature ϑG,max1)

180 °C

140 °C

140 °C

Max. Cooling Capacity Qmax

450 kJ/h

230 kJ/h

120 kJ/h

Gas Pressure pmax

160 bar

3 bar

3 bar

Pressure drop Δp (v=150 L/h)

8 mbar

8 mbar

8 mbar

Dead Volume Vtot

69 ml

48 ml

129 ml

Gas Connections (metric)

G1/4

GL14 (6mm)4)

DN 4/6

Gas Connections (US)

NPT 1/4″

GL14(1/4″)4)

1/4″-1/6″

Condensate Out Connections(metric)

G3/8

GL25(12mm)4)

G3/8

Condensate Out Connections (US)

NPT 3/8″

GL25(1/2″)4)

NPT 3/8″

Heat Exchanger

DTS(DTS-63)) DTSI(DTS6-I3))2)

DTG DTG

DTV3)DTV-I2)3)

Version / Material

Stainless steel

Glass

PVDF

Flow Rate vmax 1)

2 x 250 L/h

2 x 140 L/h

2 x 115 L/h

Inlet Dew Point τe,max1)

80 °C

65 °C

65 °C

Gas Inlet Temperature ϑG,max1)

180 °C

140 °C

140 °C

Max. Cooling Capacity Qmax

450 kJ/h

230 kJ/h

185 kJ/h

Gas Pressure pmax

25 bar

3 bar

2 bar

Pressure drop Δp (v=150 L/h)

5 mbar each

5 mbar each

15 mbar each

Dead Volume Vtot

28 / 25 ml

28 / 25 ml

21 / 21 ml

Gas Connections (metric)

6 mm tube

GL14(6mm) 4)

DN 4/6

Gas Connections (US)

1/4″ tube

GL14(1/4″)4)

1/4″-1/6″

Condensate Out Connections(metric)

Tube 10 mm (6 mm)

GL18(10mm)4)

DN 5/8

Condensate Out Connections (US)

Tube 3/8″ (1/4″)

GL18(3/8″)4)

3/16″-5/16″

 

PT. Mora Anugerah Berkat [M.A.B] adalah perusahaan yang sudah berpengalaman dalam menyediakan/mensupply peralatan laboratorium untuk gas analyzer seperti BUHLER – Gas Cooler Series RC 1.1 , kami juga menyediakan berbagai brand/merk internasional dari USA, Eropa dan China,produk internasional yang kami supply/tawarkan seperti : Buhle Tchnology, ECOM J2KN, Bacharach dan lain – lain ; Kami juga menyediakan Jasa Perbaikan (SERVICE), Jasa Perawatan (Maintenace) dan support untuk KALIBRASI / AKSESSORIES SPAREPART jangan ragu untuk menghubungi kami marketing dan sales kami siap membantu menjawab segala kebutuhan anda.

BUHLER – Gas Cooler Series RC 1.1

 

PT. Mora Anugerah Berkat (M.A.B)
Click Here : Sample Gas Cooler RC 1.1 Brochure
Click Here : Gas Cooler Series TC-MIDI Brochure
Alamat : Jl.Taman Asri Blok M No.36 Kel.Cipadu Kec.Larangan,Kota Tangerang – Provinsi Banten
Phone : 021 7345.7064
Mobile: 0812.8140.0749 / 0812.8770.5470
Chat : Whats App,Line,WeChat,Telegram
Website : www.moraanugerah.idwww.moraanugerah.com
Email : info@moraanugerah.com – marketing@moraanugerah.com – marketing_div@moraanugerah.com