Mercury Analyzer RA-50 Specification

Mercury Analyzer RA-50 Specification

 

 

 

 

Mercury Analyzer RA-50 Specification

 

The RA-5A delivers ultra-high sensitivity and excellent precision while only requiring 5ml os sample. The RA-5A is also environmentally-friendly thorughits direct purge design that reduces reagent consumption and waste production by over 80% versus comparable model without any need for compressed gases

RA-5A has achieved the highest level sensivity and accuracy with new innovatory-crafted detector optics

Sample

N

Average

C.V

0.05μg/L

5

0.052μg/L

2.2 %

0.10µg/L

5

0.101μg/L

1.70%

 

Click Here : Mercury Analyzer RA-5A Brochure 

 

Mercury Analyzer RA-50 Specification

Nippon Instruments Corporation

Specifications

Mercury Analyzer RA – 5A

Data

Processing

Software

Measurement
Principle

Non-dispersive double beam cold vaporatomatic absorption
spectrometry [Patent Pending]

Measurement
Method

Reducing Vaporrization with open air-supply
(Peak height measuring and integration measuring salectable)

Light
Source

Low pressure mercury discharge lamp

Wave
Length

253.7 nm

Detector

semicondutor sensor (w/254nm bandpass filter)

Detection
Limit

0,0025ng/5ml (0.5ppt)

Maximum
Range

2.000ng/5ml (400 ppb)

Carrier Gas
Required

No0 Compressed purified gas Required

Measure
Flow Rate

Adjustable (0.1 – 0.9 L/min) (just operet with ambient air)

Measuring
Time

30sec – 999sec

Dehumidifier

Peltier device

Mercury
Removal

Removal by activated carbon

Mercury

Detection

Section

OS

Windows 10 Pro

Comunication

USB

Display

Peak Wave shape,calibration curve,measuring time and measuring result

Data

Selectable from 6 formula (3 liniear and 3rd order fits), Correlation

Processing

coefficient is automatically calculated Static calculation : Mean Value,
standard deviation Coefficient of variation

Software

(All value are automatically calculated), self check function

Print

Measuring conditions,Calibration Curve,Measured values,Statical
Calculations Peak wave shape and Measuring time

Quality
Control

Check Standard, Avoid abnormally high mercury

Instalation
Dimension

248W x 253D x 180H (mm)

Weight

4,7 (Kg)

Instalation
Enviroment

10°C – 35°C, < 10 – 80% RH (Non condensing enviroment is allowed)

Power Supply

AC, single phase, 100 – 240V 50/60 Hz, 150VA (AC Adaptor Included)

Kami juga menyediakan Jasa Perbaikan (SERVICE), Jasa Perawatan (Maintenace) dan support untuk KALIBRASI / AKSESSORIES SPAREPART jangan ragu untuk menghubungi kami marketing dan sales kami siap membantu menjawab segala kebutuhan anda.

Mercury Analyzer RA-50 Specification

  • PT. Mora Anugerah Berkat (M.A.B)

  • Click Here : Mercury Analyzer RA-5A Brochure 

  • Alamat : Jl.Taman Asri Blok M No.36 Kel.Cipadu Kec.Larangan,Kota Tangerang – Provinsi Banten

  • Phone : 021 7345.7064

  • Mobile : 0812.8140.0749 / 0812.8770.5470

  • Chat : Whats App,Line,WeChat,Telegram

  • Website : www.moraanugerah.idwww.moraanugerah.com

  • Email : info@moraanugerah.com – marketing@moraanugerah.com – marketing_div@moraanugerah.com